Product Introduction
F72MB-DCTB Temperature Gauge with Transmitter and Alarm Output Features: Input Signal: PT100, K-type thermocouple Accuracy: 1.5 Ambient Temperature: -25℃ to +55℃ Enclosure Protection: IP44 (IP65 with protective cover; must be specified in order). Vibration Resistance: 2-100Hz, ≤0.7g Withstand Voltage: AC2.5kV, 50Hz, 1min Operating Power Supply: 24Vdc±25%, 100Vac, 220Vac+20% Power Consumption: ≤2VA Output Contact Capacity: AC250V, 1A, DC28V, 1A Alarm Point Setting: Can be configured with upper limit, lower limit, and peak lower limit according to user requirements. Measurement range is the same as F96-DCT.
Specifications
F72MB-DCTB Temperature Gauge with Transmitter and Alarm Output
Features:
Input Signal: PT100, K-type thermocouple
Accuracy: 1.5
Ambient Temperature: -25℃ to +55℃
Enclosure Protection: IP44 (IP65 with protective cover; must be specified in order).
Vibration Resistance: 2-100Hz, ≤0.7g
Voltage Withstand: AC2.5kV, 50Hz, 1min
Operating Power Supply: 24Vdc±25%, 100Vac, 220Vac+20%
Power Consumption: ≤2VA
Output Contact Capacity: AC250V, 1A, DC28V, 1A
Alarm Point Setting: Can be configured with upper limit, lower limit, and peak lower limit according to user requirements.
Measurement range is the same as F96-DCT.
Product Category:
Model
Expansion Angle
External Dimensions (A*B*C)
Opening Size (mm)
Weight (kg)
F72-DCTB
90°
72*72*120
68*68+0.5
0.3
F96-DCTB
96*96*130
92*92+0.5
0.4
F144-DCTB
144*144*130
138*138+0.5
0.45
F72B-DCT
240°
72*72*120
68*68+0.5
0.5
F96B-DCT
96*96*130
92*92+0.5
0.5
F144 B-DTC
144*144*130
138*138+0.5
0.6
Note:
Limit instruments not listed in the specification sheet can also be supplied. Users must specify this when ordering and confirm it in the contract.
See Figure 10.2 for instrument terminal block arrangement.
Users should specify the model, specifications, power supply, alarm values, and other parameters when ordering, and specify and confirm these in the contract.
For special purpose instruments, please add "J" after the first half of the model number (only applicable to 240° instruments), such as F96JB-DCT.